banner_new
Search - Communicator
Search - Jfcategories
Search - Jfcontacts
Search - Jfcontent
Search - Jfnewsfeeds
Search - Jfsections
Search - Jfweblinks
Search - Content
Search - Weblinks
Search - Contacts
Search - Categories
Search - Sections
Search - Newsfeeds
thl_eksipiretisi

Σύνδεση

We have 7 guests online
banner_m-d-s2a

There are no translations available.

Gima MB122

 • Minimum preselectable power: 0
• Level step: 1
• Maximum output power CUT (W): 120 - 250 ohm
• Maximum output power BLEND (W): 90 - 200 ohm
• Maximum output power COAG FORCED (W): 80 - 150 ohm
• Maximum output power COAC SOFT (W): 60 - 100 ohm
• Maximum output power BIPOLAR (W): 40 - 100 ohm
• Crest Factor CUT: 1.5
• Crest Factor BLEND: 2.1
• Crest Factor COAG FORCED: 2.0
• Crest Factor COAG SOFT: 1.7
• Crest Factor BIPOLAR: 1.5
• Working frequency: 600 kHz
• Weight: 7 Kg
• Size LxHxD mm: 260 x 110 x 265
• Operating voltage (switchable): 115-230
• Mains frequency (Hz): 50-60
• Fuses 230Vac (5x20) TIMED: 2x 3.15A
• Fuses 115Vac (5x20) TIMED: 2x 6.3A
• Electrical input power (VA): 300
• Electrical input current (230Vac) (A): 1.3
• Electrical input current (115Vac) (A): 2.6
• Five steps adjustable sound level (from 55- to 75dBA):
• Self-check:
• Power accuracy output warning
• Split or not split patient plate allowed
• Last working condition storing
• Electrical Class (EN60601 - 1): I CF
• MDD 93/42/EEC Class: II B
• Patient circuit: F
• Conform to EN60601-1-4 (1998)